183
19. Data Output
Data output mode
The timing of data output of the balance can be changed by using the [Data output mode] screen in the
function table.
You can configure the [Key mode], [Auto print mode], [Interval output mode] and [Stream mode] settings
in the following screen.
MENU key
System settings button
Communication button
Data output button
Data output mode button
[Data output mode] screen
Key mode
Output only when the display is stable
If the PRINT button
is pressed when the stabilization indicator is displayed, the weighing value
will be output once.
At this time, the weighing value display will blink once to show that it has been output.
Output regardless of whether the display is stable or unstable
Regardless of whether the stabilization indicator is turned on or not, the weighing value will be output
when the PRINT button
is pressed.
Output after the display becomes stable
If the PRINT button
is pressed when the stabilization indicator is displayed, the weighing value
will be output once.
If the stabilization indicator is not displayed, the weighing value will be output once the next time the
stabilization indicator is turned on.
At this time, the weighing value display will blink once to show that it has been output.
Auto print mode
Zero point
If the weighing value exceeds the range specified with [Bandwidth], [Standard] or [Polarity] in the [Data
output mode] screen in the function table from the standard "zero display" and the stabilization indicator
is displayed, the weighing value will be output once. In addition, if the PRINT button
is pressed
when the stabilization indicator is displayed, the weighing value will be output once.
At this time, the weighing value display will blink once to show that it has been output.
Last stable value
If the weighing value does not fall below the range specified with [Bandwidth], [Standard] or [Polarity] in
the [Data output mode] screen in the function table from the standard "latest value with the stabilization
indicator displayed" and the stabilization indicator is displayed, the weighing value will be output once.
In addition, if the PRINT button
is pressed when the stabilization indicator is displayed, the
weighing value will be output once. At this time, the weighing value display will blink once to show that it
has been output.
Summary of Contents for BA-T Series
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Page 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Page 117: ...117 Output example for periodic check output results Periodic check output results...
Page 118: ...118 Periodic check results PDF...
Page 129: ...129 AND MEET analysis graph example...
Page 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...