Commentary on filter settings
Response characteristics
The display shows an acute response to weight variation. Reduce the setting value when
weighing powder or liquid, weighing extremely lightweight samples, or prioritizing the
operating efficiency over the stability of the weighing value.
The display shows a gradual response to weight variation. Increase the setting value when
the weighing value does not get stabilized depending on the usage environment.
Stability bandwidth
The setting to determine that the weighing value is stable. It shows the stabilization indicator when the
fluctuation range in a certain time of period gets lower than the setting value, and outputs the weighing
value based on the internal settings. This setting affects the automatic printing.
The minimum display to be shown is 1 d.
(Example) If No blank is selected with BA-6TE and 0.000001-g display is chosen, 0.000001 g is
shown as 1 d.
The stabilization indicator does not appear unless the weighing value is stable enough, and
it disappears when the weighing values changes even slightly. Reduce the setting value for
precise weighing.
The display does not easily respond to a slight weight variation. Increase the setting value
when the weighing value does not get stabilized depending on the usage environment.
Zero tracking
The function to automatically track the zero point and keep the zero display when the display shows
zero and the zero point slightly fluctuates depending on usage environment. A tracking level can be
selected from 3 levels. Increase the setting value if zero is not stabilized. Do not perform zero tracking
when the weighing value is slightly over d. d is the minimum unit of the display.
Setting
value
BA-6TE /
BA-6DTE
BA-225TE /
BA-225DTE /
BA-125DTE
Description
Off
Off
The zero point is not tracked with zero tracking.
Normal
±2 d/1 second
±1 d/1 second
The zero point is normally tracked with zero
tracking.
Strong
±5 d/0.5 seconds ±1.5 d/0.5 seconds
The zero point is strongly tracked with zero
tracking.
Very
strong
±10 d/0.2 seconds ±1.5 d/0.2 seconds
The zero point is very strongly tracked with zero
tracking.
FAST
SLOW
±1d
±3d
Summary of Contents for BA-T Series
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Page 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Page 117: ...117 Output example for periodic check output results Periodic check output results...
Page 118: ...118 Periodic check results PDF...
Page 129: ...129 AND MEET analysis graph example...
Page 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Page 219: ......