98
Procedure for external sensitivity adjustment
Weighing example for BA-6DTE
Step 1. Make sure that nothing is on the weighing pan
and press the Confirm button
.
The balance measures the zero point.
Do not apply vibration and the like to the balance.
*
You can input the external weight value before input.
Step 2. Place the weight on the weighing pan and press
the Confirm button
.
Measure the weight.
Do not apply vibration and the like to the
balance.
*
You can input the external weight value before input.
Step 3. The [Sensitivity adjustment result] screen for the
external sensitivity adjustment is automatically
displayed. Please remove the weight.
Step 4. You can output the result by pressing the
PRINT button
.
*
The result is automatically output if [GLP output] in
[Data output] is set to ON.
Weighing pan
Confirm button
External weight
Confirm button
Result display
GLP output
PRINT button
Step 5. Press the HOME key
to return to the weighing screen.
Place the weight again to confirm that the sensitivity of the balance is adjusted properly.
If it is not within the range, start over from the first step of this procedure in the appropriate
ambient conditions.
Summary of Contents for BA-T Series
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Page 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Page 117: ...117 Output example for periodic check output results Periodic check output results...
Page 118: ...118 Periodic check results PDF...
Page 129: ...129 AND MEET analysis graph example...
Page 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Page 219: ......