155
USB interface
Display settings: MENU key
System settings button
Communication button
USB interface button
[USB interface] screen
1
USB device mode
2
USB data format
3
Format the USB flash drive button
4
Back button
Name
Setting value
(setting range)
Description
1
USB device mode
Quick USB ,
USB Virtual COM
Selects the connection method when
the USB cable is used.
2
Data format
A&D standard format ,
DP format,
KF format,
MT format,
NU format,
NU2 format,
CSV format
UFC format
You can select the data format.
3
Format the USB flash drive button
-
Format the USB flash drive.
4
Back button
-
Returns to the previous screen.
Settings in the red box are default values (factory settings).
To connect the balance with your personal computer, you can select from the Quick USB mode and
USB Virtual COM mode. For details, refer to “
18. Connecting to a Personal Computer
For specifications of the USB interface, refer to “
For details on the data format, refer to “