188
CSV format
The data part and unit part of the A&D standard format are separated by a separator ",".
Outputs the unit even when overloaded.
When the decimal comma (,) is set, a semicolon (;) will be used instead as a separator.
S T
,
+ 0 1
.
2 3 4 5 6 7
,
g
C
R
L
F
Header
Data
Unit
Terminator
S T
When stable
C
R
:
Carriage return,
ASCII 0Dh
U S
When unstable
L
F
:
Line feed,
ASCII 0Ah
Q T
Counting mode when stable
:
Space,
ASCII 20h
O L
When overloaded
If the gross/tare weight is added in the [Data to be added] screen of the function table, the second
header corresponding to the output data is appended to the header.
S T
,
N
,
+ 0 1
.
2 3 4 5 6 7
,
g
C
R
L
F
2nd header
Data
Unit
Terminator
N
Net weight
C
R
:
Carriage return,
ASCII 0Dh
G
Gross weight
L
F
:
Line feed,
ASCII 0Ah
T
Tare weight
:
Space,
ASCII 20h
P T
Preset tare weight
Display settings: MENU key
System settings button
Communication button
Data output button
Data to be added button
[Data to be added] screen
When other data is added to the weighing value, all data will be output in one line.
The output sample will be as follows if the ID number, data number, date and time are added.
SAMPLE-0123-4,No,012,2020/07/01,12:34:56,ST,+01.234567,
g
ID number Data number
Date
Time
Weighing value
TAB format
This is a format, in which the separator of the CSV format is changed from comma to TAB.
Used when connecting to a personal computer and inputting to Excel and the like.
S T
TAB
+ 0 1
.
2 3 4 5 6 7
TAB
g
C
R
L
F
Header
Data
Unit
Terminator
S T
When stable
C
R
:
Carriage return,
ASCII 0Dh
U S
When unstable
L
F
:
Line feed,
ASCII 0Ah
Q T
Counting mode when stable
:
Space,
ASCII 20h
O L
When overloaded
TAB
:
Horizontal tab,
ASCII 09h
Summary of Contents for BA-T Series
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Page 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Page 117: ...117 Output example for periodic check output results Periodic check output results...
Page 118: ...118 Periodic check results PDF...
Page 129: ...129 AND MEET analysis graph example...
Page 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Page 219: ......