14
BA-6TE / BA-6DTE
BA-225TE / BA-225DTE / BA-125DTE
Installation considerations, preparation and precautions
Prepare the following installation conditions in order to bring out the full performance of the balance.
Note that the installation environment needs to be taken into consideration for the highly sensitive BA-
6TE / BA-6DTE.
Install the balance in an environment where the temperature and relative humidity are not excessive.
The best operating temperature is about 20°C ±2°C at about 45% to 60% relative humidity.
Install the balance where it is free of dust.
The weighing table should be solid. (An anti-vibration table or stone table is ideal)
We recommend using an anti-vibration table (AD-1671) for BA-6TE.
It is advisable to install the display unit in a location other than on the anti-vibration table so that errors
due to tilt that occurs when operating the switches of the balance on the anti-vibration table can be
avoided.
Place the balance on a horizontal table, and make sure that it is not tilted.
Install the balance in a stable location, avoiding vibration and shock. Corners of rooms on the first floor
are best, as they are less prone to vibration.
Install the balance where it is not affected by heaters or air conditioners. Avoid breezes and drafts in the
room.
You can reduce the influence of breezes and drafts by using an AD-1672/AD-1672A (large size) or AD-
1676 (medium size) tabletop breeze break.
Install the balance where it is not exposed to direct sunlight.
Install the balance away from equipment which produces magnetic fields.
Level the weighing unit with the leveling feet and bubble spirit level. Refer to “
Summary of Contents for BA-T Series
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Page 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Page 117: ...117 Output example for periodic check output results Periodic check output results...
Page 118: ...118 Periodic check results PDF...
Page 129: ...129 AND MEET analysis graph example...
Page 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Page 219: ......