Summary of Contents for ZS Series

Page 1: ......

Page 2: ...1 INDEX Installation 2 Specifications 3 Key Layout Functions 4 Internal Functions 6 Initial Setup 8 Instructions for Use 11 Error Codes Troubleshooting 15 Daily Care Maintenance 17 ...

Page 3: ...e external power adaptor Before plugging the power adaptor into the electricity outlet check and make sure the input voltage of the power adaptor matches with the output voltage of the outlet If do not match please do not plug in the adaptor and contact your dealer When the power adaptor is being used please remove all the batteries from the battery compartment Failure to do so might cause a batte...

Page 4: ...ernal Power Adaptor Max Zero Range 2 of Rated capacity Max Tare Range 100 of Rated capacity Subtractive Weight Units Note A Metric g Avoirdupois lb oz Taiwanese tg Operation Environment 0o 40oC 32o 104oF Non condensed R H 85 Specifications subject to change without notice NOTE A To comply with the law of certain countries Avoirdupois and Taiwanese weight units may be disabled Contact your dealer f...

Page 5: ... OFF KEY Press this key to turn scale off ON ZERO KEY Press this key to turn scale on or set weight to zero TARE KEY Press this key to tare off the weight of a container MODE KEY Press this key to switch between weight units ...

Page 6: ...ppears recharge the scale immediately WEIGHT INDICATOR Displays the weight of the item placed on the scale NET INDICATOR This indicator appears to indicate the tare function is in operation and weight shown is net weight METRIC INDICATOR This indicator appears to indicate the metric weight unit is in use AVOIRDUPOIS UNITS INDICATOR This indicator appears to indicate the avoirdupois weight unit is ...

Page 7: ...s Fun 4 Select Auto Power Off Fun 5 Select Rs 232 Baud Rate Protocol Fun 6 Select Digital Motion Filtering Function Fun 7 Select Serial Output Mode Fun 8 Select UTP Printout Format This function is UTP related Fun 9 Select Auto Tare Function Fun 12 Select Number Of Printout Copy This function is UTP related Fun 15 Set and Send Date Time Data to UTP This function is UTP related Fun 19 For Qualified...

Page 8: ... have been reserved for future use they are not functional Do not alter any parameters set in Fun 1 19 which may cause system unstable error or non accurate weighing result If entered by mistake press OFF immediately Contact your dealer for more information ...

Page 9: ...at d e for next other setting or i Press TARE to save and restart scale or j Press OFF to save without restarting Display Segment Check Under this function backlight if equipped functional indicators and display segments will be lit up Contact your dealer if any display element is missing Select Auxiliary Weight Units 2 auxiliary weight units in addition to metric are available These are Avoirdupo...

Page 10: ...Avoirdupois and Taiwanese weight units all enable Select Auto Power Off Function 5 parameters are available 0oFF Auto Power Off Function disable 4oFF Auto off when left idle for 4 min 10oFF Auto off when left idle for 10 min 20oFF Auto off when left idle for 20 min 30oFF Auto off when left idle for 30 min Under the Auto Power Off function mode the scale will turn off automatically when it is left ...

Page 11: ... is subject to disturbances like vibration wind flow etc Select higher filter values for worse environments Filt2 gives the slowest response compared with other 2 parameters Select Auto Tare Function 2 parameters are available Tr On Auto Tare Function enable Tr OFF Auto Tare Function disable Under the Auto Tare Function the first load applied to scale will be treated as a container and will be tar...

Page 12: ...le is turned on 4 The ZERO INDICATOR is on If not press ON ZERO to set display to zero Weighing 1 Always place an object onto the platter gently Excessive force applied to platter may cause damages to sensor inside scale 2 The weight of the object would be displayed automatically 3 It is a good practice to remove all loads from platter after weighing It would prolong the lifetime of scale ...

Page 13: ...gain Set Display Value to Zero when Unloaded Press ON ZERO to set weight displayed to zero when unloaded Refer to SPECIFICATIONS for maximum zero range Tare off the Weight of a Container When a container is used follow the below steps to tare off the weight of it and to obtain a net weight result a Remove all loads away from platter b Make sure that zero weight and the ZERO INDICATOR appear If not...

Page 14: ... page 11 for weighing procedures Weight displayed is the net weight of the subject matter Refer to SPECIFICATIONS for maximum tare range Clear the Tare Function a Remove all loads from platter b The scale will appear a negative value which is the weight of the container and the ZERO INDICATOR appears c Press TARE d After TARE is pressed a zero weight will be displayed while the ZERO INDICATOR rema...

Page 15: ...means that the battery is faded out Replace all battery with new ones or apply the external power adaptor immediately Caution When the power adaptor is being used remove all batteries from the battery pack Failure to do so may cause battery leak explosion or damages to the scale ...

Page 16: ...urned on Check Is the scale powered or battery faded out Action Replace the battery with a new one or apply the power adaptor Check Is the power adaptor inserted properly into both the electricity outlet and the DC inlet of scale Action Secure both ends of the power adaptor and try again Syndrome Scale displays Err Check Is the load applied exceeding the rated capacity of scale Action Remove all l...

Page 17: ...y with a new one or apply the power adaptor immediately Syndrome Weighing result is not accurate Check Is the scale placed in a level condition Action Relocate the scale to obtain a level condition Check Is the scale affected by airflow vibration or RFI Action Place the scale away from all disturbances Check Is the scale calibrated correctly Action Contact your dealer Check Is the correct weight u...

Page 18: ...detergent 2 Do not use any harsh abrasive material acetone volatile solvent thinner or alcohol for cleaning 3 Verify the accuracy of scale periodically Re calibrate the scale if necessary 4 Remove weights from the platter when not in use 5 Recharge the battery regularly when the scale is not in use ...