NOTE
It is recommended that the “Preview” button be activated; the user can decide whether the alignment has been
successful using the preview.
Examples
The areas shown in green in the left-hand image (with Preview active) show the correspondence of the alignment.
The user can decide whether the alignment is satisfactory or whether new or further alignment points need to be defined.
The result of the alignment can be accepted by clicking
”OK”
1
or a more accurate automatic precision alignment can be
carried out using the
“Align”
2
button.
4.5.9.3 Application of the Align function
In order to be able to use the automatic precision alignment, it is first necessary to mark three identical points
→
Chapter 4.5.9.2 “Manual three-point alignment”. The precision alignment of the two data sets is carried out with reference
to their object geometry based on this preliminary alignment (best-fit alignment).
Align function settings
Align function dialog box Information relating to the result of the precision alignment process
1 Start
The alignment of the data sets is optimized based on the set parameters.
2 Subsampling
Sampling rate. This rate determines how many data points are to be used for optimizing the precision alignment –
each data point (1 / 1) or fewer data points.
3 Search Range
The search range defines the radius of the overlapping area within which corresponding data points are sought.