35
4 Max. iterations
Determines the maximum number of iterative optimization loops.
5 Only Selection
NO:
the complete data set is used for precision alignment
YES:
only pre-selected data areas are used for precision alignment.
6 3D View Update
NO:
3D image is not updated (less memory usage)
YES:
Every change in an optimization loop is updated in the 3D image
NOTE
Once changed, all parameter settings (Subsampling, Search Range etc.) are used as the default value for other projects.
4.5.10 Match holders (markers) / Scan bodies
The coded match holders are a special feature, as these enable the
position of the implants to be determined unambiguously.
Using the coded match holders, a file is created for each tooth;
angle and coordinate data are accurately measured and transferred
to Zfx Manager 2.0 for further processing.
The uncoded match holders are used as scan bodies and are only
measured as such.
CAUTION
Do not modify or contaminate the codes on match holders or render them unrecognizable, as this can prevent the
markers from being recognized correctly.
The match holders must be properly recorded by the scan. If this is not possible due to their close arrangement,
a marker scan should be carried out with the Add-Scan function (in the working step).
Coded markers can be used several times per restoration.
4.5.10.1 Match holder (marker) coding
The match holders are coded by Zfx GmbH and supplied as an optional accessory.
NOTE
It is recommended that only coded match holders be used, as this is the only way to avoid extensive rework in the
Zfx Design CAD software.