Zfx™ Evolution
NEXT
30
4.5.6.3 Possible problems with the additional scan
If some of the defective areas are not closed in spite of the additional scan, it may be necessary to treat the scan object
with an anti-reflection spray. This is, however, only recommended for strongly reflective (shiny) or partly transparent
(impression materials) surfaces.
CAUTION
Do not spray anti-reflection spray into the scanning chamber. Always spray the scan object outside the scanning
chamber as otherwise there is a risk of the system optics becoming contaminated.
4.5.6.4 Starting an additional scan for already completed models
An additional scan is only possible while in Operating Mode. If it is necessary to gather data for already completed models,
the appropriate scan must be restarted from the required point.
NOTE
Only complete the scanning process after carefully checking whether sufficient scan data have been gathered.
4.5.7 Additional Scan – Add-Scan
A scan order with, for example, three specified scan steps (jaw model / teeth / mush bite)
can be extended by one or more scanning steps using the “Add-Scan” function.
This function is necessary when further information relating to a restoration is required
or the scanned data do not cover the desired area.
4.5.7.1 Start Add-Scan
×
Click the
“Add-Scan”
button
×
Position the scan object in the scanner
NOTE
As the scanning step is not defined in the Dental Manager or via the configurator, no preview is available using the
Scan Assistant. The additional scanning step is also displayed as an Add-Scan image after the scan.
×
Start Add-Scan by clicking the
“Scan”
1
button.
×
Enter the name and setting and confirm by clicking
“OK”
2
2
1
Содержание Zfx EvolutionNEXT
Страница 1: ...Zfx EvolutionNEXT Manual A COMPANY OF ...
Страница 7: ...7 ...
Страница 13: ...13 ...
Страница 45: ...45 ...
Страница 65: ...65 ...