IOLMaster 700
3.2.5
Dust Covers
Precondition:
The measurement head side covers have been removed.
Procedure:
1)
Remove the front dust cover
2)
Disconnect the power cable of the cooling
fan and remove the side dust cover
(
3)
Remove the screw (Torx 10) and
disconnect the cable of the temperature
sensor. Lift up the OCT Engine 1060
(
).
4)
Remove the top dust cover (
Reassembly:
1)
Reassemble in the reverse order.
Fig. 22 - Side and Front Dust Cover
Fig. 23 - Top Dust Cover
Side
Front
Screw
Top
Service Manual
Confidential and Proprietary
SM-30-4060
Order No. 2075-325
3 - 10