IOLMaster 700
4.4.2
Paddle Polarisation
The Paddle Polarisation procedure allows accurate adjustment of the paddle position to ensure
maximum SNR during measurement. The paddle polarisation tool will be automatically removed after
restarting the instrument, therefore it must be installed each time the procedure is performed.
Procedure:
Enter the service mode
1)
Ensure the File Based Write Filter protection
2)
is ENABLED.
Fig. 44 - File Based Write Filter Enabled
Using Windows Explorer, navigate to the
3)
folder: Y:\PolarizationToolFiles.
Run (double click) the
4)
uEye_LabView_40200.exe.
Fig. 45 - Run the uEye_LabView_40200.exe
Select “Englisch” and press “weiter >”.
5)
Press “Next >”, press “Next >” and press
6)
“Next >”.
Press “Install”.
7)
Fig. 46 – Install the file
1
2
Service Manual
Confidential and Proprietary
SM-30-4060
Order No. 2075-325
4 - 20