IOLMaster 700
Instructions: (Step 2)
(1) Wait, till output-window indicates that
work is done.
(2) Check the ‘Console-Output’ for warnings
and/or error messages.
(3) Click 'Finish' button.
Fig. 16 – S301 Exchange complete measurement head
(3)
Attention!
The factory-adjusted device-settings must be verified against the local test eye.
For this reason, it’s recommended to close the Service Tool, restart the device and
complete a calibration check using the IOLMaster 700 application!
S401 – Exchange complete computer base with OS & application pre-installed
Description:
Use this function after the computer base
has been replaced and both disk-image and
user-data have been restored.
This use-case completes the following tasks:
(1) Set the serial-number of the new
computer base.
(2) Retrieve device-settings from both
measurement head and local files.
(3) Join both sets of device-settings:
Use serial-number of computer base from
manual input,
use all other device-settings from
measurement-head (mirror).
(4) Persist the joined settings to both
measurement-head and local-files.
(5) Backup the joined settings onto the
service-drive (mSata).
Fig. 17 – Exchange complete computer base with OS &
application pre-installed (1)
Service Manual
Confidential and Proprietary
SM-30-4060
Order No. 2075-325
4 - 10