IOLMaster 700
BACKUP (mSATA SSD 32 GB)
Bootable drive containing the ‘Acronis True Image Personal’ application for disk backup/recovery.
Paddle Polarisation tool files
Y:\PolasizationToolFiles\ (See section
for further details).
Config files
(backed up after using the Service Tool):
Y:\DeviceConfiguration_Backup\ (config, device.settings.xml, filterGenerator.settings.xml)
Acronis Images:
Y:\Factory Image\*.tib (factory image of System and Data drives)
STM32_PUB (EEPROM 68KB)
Windows can not be used for copy/cut/paste of files into the measurement head (STM_PUB32).
Measurement head configuration files can be found on both the computer base (System) and the
measurement head (STM_PUB32) – they must be identical (this can be confirmed with the service
tool)!
Z:\config (binary configuration file = C:\DeviceConfiguration\config)
non-visible/accessible: Z:\device.settings.xml
non-visible/accessible: Z:\filerGeneartor.settings.xml
Service Manual
Confidential and Proprietary
SM-30-4060
Order No. 2075-325
4 - 28