![background image](http://html1.mh-extra.com/html/zeiss/iolmaster-700/iolmaster-700_service-manual_3344775020.webp)
Parts Removal Replacement
IOLMaster 700
3)
There are two variations of fasteners installed
in the side cover.
Old – only a quarter turn, anticlockwise
See Fig. 3
New – several turns anticlockwise
See Fig. 4
Fig. 3 - Old fastener, quarter turn
Fig. 4 - New fastener, several turns
4)
Remove the left side cover.
5)
Remove the right side cover by loosening the
four holding screws (Torx 20 – 100mm) as
shown in
Fig. 5
.
Note
Care should be taken to avoid damaging
sensitive parts (lenses) whilst gaining
access to these screws.
Reassembly:
1)
Reassemble in the reverse order;
Old – the fasteners on the right side cover
should be set to the closed position (
Fig. 3
)
before attaching.
New – the fasteners must be open (
Fig. 4
)
before attaching.
Fig. 5 - Right cover
Caution!
When performing service activities with
measurement head covers off. Particular
attention should be given to not damage
or move the polarisation paddles.
They can’t be re-adjusted in the field!
Fig. 6 - Polarisation paddles
Screws
Open
Closed
Closed
Open
Warning!