Instructions: (Step 1)
Prerequisites:
–
The old PC has been replaced by a new
one.
–
Either:
the old hard drive has been installed in
the new computer base
OR
the new hard drive must be properly
prepared (Acronis-image installed, user-
data restored, etc.)
–
Measurement head and PI were not
changed.
–
The computer base OS is running
without errors.
–
The 'C:\DeviceConfiguration' contains
the local device-settings files (default
values (new hard drive) or values from
old computer base (old hard drive)).
(1) Ensure all prerequisites are fulfilled.
(2) Enter the serial-number of the new
computer base. This is a numeric ID,
similar to '1102301' (labelled serial no.).
(3) Click button 'Continue' to proceed
or ‘Step Back’ to the use-case selection
view.
Fig. 18 – Exchange complete computer base with OS &
application pre-installed (2)
Fig. 19 – Computer Base Serial Number
Instructions: (Step 2)
(1) Wait, till output-window indicates that
work is done.
(2) Check the ‘Console-Output’ for warnings
and/or error messages.
(3) Click 'Finish' button.
Fig. 20 – Exchange complete computer base with OS &
application pre-installed (3)
SM-30-4060
Confidential and Proprietary
Service Manual
4 - 11
Order No. 2075-325