4.2
Service Mode
The IOLMaster 700 has a service mode which can be accessed using the Windows user account:
CZMService. The passwords must not be distributed to any non-authorised person.
The usage of the service mode is restricted to certified field service engineers (FSEs), trained from an
authorised ZEISS trainer.
4.2.1
Entering Service Mode
The service mode can be accessed in three
different ways:
Method 1 (no keyboard required):
Log in to the application software as user
1)
“Service” (for details about changing user
refer to the on-screen help “?”)
Fig. 1 - Service user application software
Enter the Settings menu and navigate to
2)
“Service”
Click “Start service mode”.
3)
Proceed to Step 8).
4)
Fig. 2 - service mode settings menu
Method 2 (keyboard required)
In any screen of the application software
5)
press the keyboard combination
“Ctrl, Alt + Del” and select “Log off”.
Proceed to Step 8).
6)
Fig. 3 – Windows Log off
SM-30-4060
Confidential and Proprietary
Service Manual
4 - 3
Order No. 2075-325