![background image](http://html1.mh-extra.com/html/zeiss/iolmaster-700/iolmaster-700_service-manual_3344775027.webp)
3.2.6
Lens Drive (PCB & Cable)
Precondition:
Shut down and switch off the instrument
The measurement head side covers have been removed from the IOLMaster 700.
The side or top dust cover has been removed.
Procedure:
Caution!
When disconnecting the connectors be
careful to not accidentally
damage/remove the socket.
1)
Disconnect the connector at main control
PCB.
2)
Remove four screws (Torx 8) on the fixation
points.
Fig. 24 - Lens Drive PCB Overview
Reassembly:
1)
Reassemble in the reverse order.
2)
Perform a system checkout in accordance
with the instrument checklist.
Fig. 25 - Lens Drive PCB
Screws
SM-30-4060
Confidential and Proprietary
Service Manual
3 - 11
Order No. 2075-325