![background image](http://html1.mh-extra.com/html/zeiss/iolmaster-700/iolmaster-700_service-manual_3344775029.webp)
3.3
Computer Base
3.3.1
Computer Base Covers
Precondition:
The instrument has been shut down and all cables have been removed.
Procedure:
1)
Remove the two screws (Torx 10) from the
cover (operator side) of the computer base.
2)
Gently remove the cover by pulling down
and away from the computer base.
3)
Remove the power fuse (refer to User manual
for detailed instructions).
4)
Place the instrument so that the side hangs
over the edge of the table. Ensure that there
is enough room to allow the side cover to be
removed without being blocked by the table.
5)
Gently pull the two tabs downwards and
remove the cover by pulling away from
computer base. A slotted screwdriver can
assist in freeing the tabs.
6)
Repeat steps 4) and 5) for the other side
cover.
Reassembly:
1)
Reassemble in the reverse order, observing
the following:
2)
When mounting the cover (operator side)
slide into the upper position paying attention
to not damage the plastic handles.
Fig. 28 – Computer Base Covers
Fig. 29 - Room for side cover removal
Screws
Tabs
SM-30-4060
Confidential and Proprietary
Service Manual
3 - 13
Order No. 2075-325