Troubleshooting
Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
M-6
Blind spot monitor reports many fixation losses • Size V blind spot check size may be too easy to see at blind spot (
).
- Consider using Gaze Track only and turn off Blind Spot.
Blue stimulus is on but bowl is not yellow
• Blue stimulus color was chosen instead of Blue-Yellow parameter (
).
- Blue Yellow parameter set to ON will give yellow bowl and Blue Size V stimulus.
Illumination from yellow lamp at top of bowl
annoying to patient
.• Blue-Yellow visor not extended during testing procedure (
Testing Problems
Test completes immediately after START button
is pressed
• SIMULATION set to ON (
- Go to Additional Setup screen, set SIMULATION to OFF.
• Instrument motorboard was out of calibration and PROCEED was selected.
Full field test stops before complete
• Trial lens holder in up position for peripheral portion of test.
- test peripheral field without a trial lens and put holder down (
Hard Disk Problems
Message indicates hard disk failure
• Refer to “How to Handle Hard Disk Failures” for step-by-step instructions (
).
Patient names exist in database without test
data
• Test results not saved with patient information.
- Use CLEANUP HARD DISK DATABASE feature to remove (
).
USB Media Problems
Message indicates “Unrecognized format” error,
USB device does not show up on the HFA, or a
USB media error is indicated.
• The USB storage device is incorrectly formatted. Format the USB storage device in
FAT (FAT16) or FAT32. NTFS or exFAT (FAT64) cannot be used.
• The USB device may be incompatible with the HFA.
Floppy Disk Problems
Floppy disk cannot be used by HFA II-
i
• Disk may not be properly formatted (
).
- Use INITIALIZE DISK feature (all previous data will be erased).
• Disk may be the wrong format (
- Make sure you use only pre-formatted High Density (HD) 1.44 MB floppy disks.
- Do not use Super High Density (2.88 MB) disks.
Floppy disk error is indicated
• Disk may be damaged or corrupted (
).
- Use REBUILD REMOVABLE MEDIA DATABASE feature (
Data Transfer Problems
Transfer of files between HFA I, HFA II and HFA II
-
i
series via serial cable does not work
• Check cable attachments (
• RS-232 port settings are incorrect (
, HFA I Manual).
Tests missing after data transfer
• Tests were ineligible to be transferred (
Table M.1 Troubleshooting Table
Problem
Possible Cause or Solution
Содержание Humphrey II-i series
Страница 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Страница 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Страница 40: ...Introduction Instrument Setup Humphrey Field Analyzer II i series User Manual 2660021145640 A 1 32...
Страница 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Страница 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Страница 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Страница 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Страница 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Страница 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Страница 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Страница 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Страница 540: ...Installing Licensing HFA II i Software Humphrey Field Analyzer II i series User Manual 2660021145640 A J 10...
Страница 558: ...Reference to Older Test Strategies Humphrey Field Analyzer II i series User Manual 2660021145640 A L 10...
Страница 579: ......