Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
Introduction & Instrument Setup
1-11
Product Compliance
0297
Complies with 93/42/EEC Medical Device Directive.
Complies with US and Canadian medical electrical system safety requirements.
Product Safety
This instrument is classified as follows:
• Class I Equipment – Protection against electrical shock.
• Type B – Degree of protection against electric shock of applied part (chinand forehead rests,
and patient button).
• Ordinary Equipment (IPX0)– Degree of protection against ingress of liquids (none).
• Continuous Operation – Mode of operation.
General Safety Requirements
• Although the HFA II-
i
is designed for continuous operation, it should be turned off when not
used for an extended period of time. The HFA II-
i
should be used in a cool, dry dust-free setting.
• Use the instrument cover to protect the HFA II-
i
at all times when it is not in use.
• Do NOT place the cover over the instrument when the HFA II-
i
is turned on, as loss of proper
airflow can cause overheating and damage to sensitive components.
• Do NOT connect or disconnect cables while power is on.
• Do NOT place any objects on top of the instrument.
• Do NOT place any container holding liquid near the instrument.
• Use only a stand or table recommended by Carl Zeiss Meditec.
Warnings and Cautions
WARNING: Do NOT block the ventilation openings. These allow for the release of
heat generated during operation. A buildup of heat due to ventilation opening
blockage can cause failures which may result in a fire hazard.
WARNING: To prevent electric shock, the instrument must be plugged into an earthed
ground outlet. Do not remove or disable the ground pin. Only an authorized Carl
Zeiss Meditec service representative may install the instrument.
WARNING: Do not use the printer or the instrument or the optional power table with
an extension cord or a power strip (multiple portable socket outlet). For additional
safety, do not plug the printer and the instrument (or the optional power table) into
the same wall outlet. Failure to observe this warning could result in electrical shock to
the patient and/or examiner.
WARNING: Do not open the instrument covers. Opening the instrument covers could
expose you to electrical and optical hazards.
Содержание Humphrey II-i series
Страница 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Страница 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Страница 40: ...Introduction Instrument Setup Humphrey Field Analyzer II i series User Manual 2660021145640 A 1 32...
Страница 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Страница 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Страница 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Страница 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Страница 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Страница 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Страница 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Страница 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Страница 540: ...Installing Licensing HFA II i Software Humphrey Field Analyzer II i series User Manual 2660021145640 A J 10...
Страница 558: ...Reference to Older Test Strategies Humphrey Field Analyzer II i series User Manual 2660021145640 A L 10...
Страница 579: ......