Unknown Networking Error Messages
In some cases the error message will not specify the possible problem.
below provides
some guidance for solving these errors.
Table I.3 Unknown Networking Error Messages
If you see
While you are
Check the following
“Error occurred archiving the
database. A network error has
occurred.”
Archiving the database.
• Ensure network cable is connected.
• Ensure Archive root folder, instrument folder and instrument's
archive folder are present on server
(<root>/<instrument>/archive/).
• Write permission set correctly on file server.
Empty HFA Selection menu.
Retrieving patient data.
• Ensure Archive root folder and instrument folder for other
instruments are present on server.
“Error occurred retrieving data. A
network error has occurred.”
Retrieving patient data.
• Ensure network cable is connected.
• Ensure Archive root folder and instrument folder for other
instruments are present on server.
“Backup to File Server failed. A
network error has occurred.”
Backing up patient database.
• Ensure network cable is connected.
• Ensure Archive root folder is present on server.
• Write permission set correctly on file server.
“Restore from File Server failed.
A network error has occurred.”
Restoring patient database.
• Ensure network cable is connected.
• Ensure Archive root folder, instrument folder and instrument's
backup folder are present on server
(<root>/<instrument>/backup/).
General Networking Error Messages
below gives some possible solutions to general error messages.
“Error occurred packaging database.”
b
• Unrecoverable write error on hard drive - replace hard drive.
“Error opening <file name>.”
• The file you are attempting to open appears to be corrupt. If it
fails to open a second time, it probably is damaged and unusable.
a. Retrieve and Restore operations only.
b. Archive and Backup operations only.
Table I.2 Specific Networking Error Messages
Possible Specific Condition
What to do
Содержание Humphrey II-i series
Страница 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Страница 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Страница 40: ...Introduction Instrument Setup Humphrey Field Analyzer II i series User Manual 2660021145640 A 1 32...
Страница 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Страница 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Страница 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Страница 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Страница 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Страница 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Страница 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Страница 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Страница 540: ...Installing Licensing HFA II i Software Humphrey Field Analyzer II i series User Manual 2660021145640 A J 10...
Страница 558: ...Reference to Older Test Strategies Humphrey Field Analyzer II i series User Manual 2660021145640 A L 10...
Страница 579: ......