Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
File Functions
10-23
On the HFA II-
i
Series:
1. From the System Setup screen, select COMMUNICATIONS SETUP.
2. Select RS-232 SETUP.
3. On the RS-232 Setup screen, set the Receive Settings for BAUD RATE to 9600, PARITY to Even,
DATA BITS to 7, and STOP BITS to 1.
4. Press PROCEED. The Communications Setup screen will then appear. Press DONE.
5. Press the
FILE FUNCTIONS
icon on the right of the screen.
6. Select TRANSFER TESTS.
7. Select CLASSIC SERIAL as the Source.
8. Choose HARD DRIVE or FLOPPY as the Destination, then press PROCEED.
On the Older HFA II:
1. From the System Setup screen, select SAVE/TRANSMIT OPTION.
2. Select RS-232 OPTIONS.
3. On the RS-232 Options screen, set BAUD RATE to 9600, PARITY to Even, DATA BITS to 7, and
STOP BITS to 1.
4. Press PROCEED.
a. For HFA II perimeters using system software version 14.0 or higher, the Save/Transmit Option
screen will appear. For Export Format, select HFA II SERIAL. Select PROCEED.
b. For HFA II perimeters using system software version lower than 14.0, you will need to update
to system software version 14.0 or later.
5. Press the
FILE FUNCTIONS
icon on the right of the screen.
6. Select TRANSFER TESTS.
7. Select HARD DRIVE or FLOPPY as the Source.
8. Choose HFA SERIAL CABLE as the Destination.
9. Choose HFA II SERIAL for the Export Format. Press PROCEED.
10. Choose the tests you wish to transfer then press PROCEED to start transferring.
On the HFA II-
i
Series:
1. A pop-up window will update you on the progress and success of test transfer on both HFA’s.
Check the HFA II to determine the end of the test transfer. It will also note how many tests
transferred. The HFA II-
i
will also indicate the number of tests successfully transferred. When
the “tests transferred” count on the pop-up window has not changed for several seconds on
the HFA II-
i
, the test transfer is complete. Press CANCEL. If you do not press CANCEL within
about 10 minutes, the HFA II-
i
will automatically proceed to the next step.
2. The transferred tests will then be moved from temporary to permanent storage. Press OK when
the pop-up window informs you that the transmission is complete.
How to Transfer Tests from One HFA II-
i
to a Second HFA II-
i
via Serial Cable
1. Connect the two HFA II-
i
instruments with an HFA II to HFA II-
i
Serial Transfer Cable (P/N
52417). Plug one end of the cable into the Data Transfer port on each of the HFA II-
i
series
instruments (refer to the diagram in
Chapter (1), "Introduction & Instrument Setup,"
port location).
2. Turn both instruments ON and wait until you see the Main Menu screen.
On each HFA II-
i
:
1. From the System Setup screen, select COMMUNICATIONS SETUP.
2. Select RS-232 SETUP.
3. On the RS-232 Setup screen, set both the Transfer Settings and the Receive Settings for BAUD
RATE to 9600, PARITY to Even, DATA BITS to 7, and STOP BITS to 1.
4. Press PROCEED. The Communications Setup screen will then appear. Press DONE.
Содержание Humphrey II-i series
Страница 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Страница 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Страница 40: ...Introduction Instrument Setup Humphrey Field Analyzer II i series User Manual 2660021145640 A 1 32...
Страница 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Страница 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Страница 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Страница 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Страница 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Страница 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Страница 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Страница 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Страница 540: ...Installing Licensing HFA II i Software Humphrey Field Analyzer II i series User Manual 2660021145640 A J 10...
Страница 558: ...Reference to Older Test Strategies Humphrey Field Analyzer II i series User Manual 2660021145640 A L 10...
Страница 579: ......