Networking
14-41
records and patient management systems use work lists to group patient data needed for particular
tasks, like exam scheduling. A work list is a list of patients specifying what services are required, or
have been provided, for them. This allows all patient data and scheduling information to be entered
in the EMR/PMS/DICOM system and then later imported into your HFA II-
i
. Work list information
might include patient names, exam types, exam dates and times. Additional data elements in the
patient database include patient last name, patient given name, patient gender, requesting
physician name, and accession number. See your EMR/PMS/DICOM documentation for more
information on using work lists.
When patient information is recalled from a work list, a new patient record is created in your HFA II-
i
database if the corresponding patient record does not exist. If the patient record exists in your HFA
II-
i
, the patient record will be updated with information from the work list. See
Rules and Conflicts,” on page 14-59
.
Exporting to EMR/PMS/DICOM Systems
To export exam data and reports to EMR/PMS/DICOM systems, use the same export functions
(Transfer Tests, Save & Transmit, and Print To a File) as you would exporting to a computer file,
except select the Transfer Destination to EMR/PMS Host for EMR/PMS systems and DICOM systems
using DICOM Gateway 1.0, or DICOM Archive for DICOM systems using DICOM Gateway 2.0.
Note: DICOM Gateway 2.0 can be set to automatically export raw exam data at the end of an exam
by enabling the END OF TEST EXAM DATA EXPORT option on the DICOM Gateway Services screen
(see
“Enable/Disable DICOM Gateway Services:,” on page H-4
Setting Up Save/Transmit for EMR/PMS/DICOM Systems
The following is a procedure to set up Save & Transmit for EMR/PMS/DICOM systems.
1
From the Main Menu, select
SYSTEM SETUP
>SAVE/TRANSMIT OPTION to display the screen
2
Open the Transfer Destination drop-down box and select EMR/PMS HOST for EMR/PMS systems
and DICOM systems using DICOM Gateway 1.0 (
), or DICOM ARCHIVE for DICOM
systems using DICOM Gateway 2.0 (
).
Содержание Humphrey II-i series
Страница 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Страница 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Страница 40: ...Introduction Instrument Setup Humphrey Field Analyzer II i series User Manual 2660021145640 A 1 32...
Страница 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Страница 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Страница 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Страница 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Страница 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Страница 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Страница 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Страница 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Страница 540: ...Installing Licensing HFA II i Software Humphrey Field Analyzer II i series User Manual 2660021145640 A J 10...
Страница 558: ...Reference to Older Test Strategies Humphrey Field Analyzer II i series User Manual 2660021145640 A L 10...
Страница 579: ......