Open the Transfer Destination drop-down box and select DATA EXPORT HOST. This selects your
PC to receive the data and act as the file server. For EMR/PMS/DICOM systems, see
to EMR/PMS/DICOM Systems,” on page 14-41
.
3
Open the Save/Transmit Option drop-down box and select SAVE AND TRANSMIT.
4
Open the Data Format drop-down box and select XML FILES or XML AND IMAGE FILES.
Use XML Files to transfer only textual information (patient data and test results) to the network
file server. You will not get a printout image with this setting. Use XML and Image Files to
transfer both graphics (a TIFF or PDF image file of the test printout) and textual information to
the network file server (see
“Printing To a File,” on page 14-7
Note: If your network server connection is lost or gives an error at the end of your exams, and your
“Save/Transmit Option” is set to “Save and Transmit”, your completed exams will still be saved to
your hard disk but not transferred to your network file server. You can choose to defer the transfer
and continue to take exams, and when the network is back up, use “Transfer Deferred Tests” to
transfer them to your file server (See
“Transfer Deferred Tests,” on page 14-34
).
5
Press PROCEED. Now when tests are saved the data will be transmitted to your file server (Data
Export Host) after the HFA II-
i
saves your completed exam to the hard drive and optionally to a
USB storage device.
Transfer Deferred Tests
If the Transfer Deferred Tests and Clear Deferred Tests buttons are active and not greyed out (
), you have completed exams that have been saved to your hard drive, but have not yet been
transferred to your network file server or EMR/PMS/DICOM system.
Figure 14.37 Save/Transmit Option screen with Active Transfer Deferred Tests button
This happens when a transfer failed due to a network error and your “Save/Transmit Option”
drop-down box is set to “Save and Transmit” in the Save/Transmit Option function. When a transfer
fails, a dialog box will be displayed to allow you to defer the transmission of the exam as shown in
. This dialog box will only be displayed for the first failed exam transfer of the session
Содержание Humphrey II-i series
Страница 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Страница 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Страница 40: ...Introduction Instrument Setup Humphrey Field Analyzer II i series User Manual 2660021145640 A 1 32...
Страница 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Страница 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Страница 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Страница 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Страница 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Страница 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Страница 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Страница 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Страница 540: ...Installing Licensing HFA II i Software Humphrey Field Analyzer II i series User Manual 2660021145640 A J 10...
Страница 558: ...Reference to Older Test Strategies Humphrey Field Analyzer II i series User Manual 2660021145640 A L 10...
Страница 579: ......