Testing
Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
5-4
The three controls on the video eye monitor are: a plus sign (+) to brighten the image, a minus sign
(-) to dim the image, and an OFF button to turn off the eye monitor display. To re-display the
monitor, press the upper-left EYE button.
Figure 5.2 The Video Eye Monitor
Gaze Tracking
(Models 740
i
, 745
i
, 750
i
)
Gaze Tracking is a unique fixation tracking system that records whether the patient is fixating
properly while stimuli are being presented. A brief initialization procedure is required at the start of
each test to calibrate and adjust the Gaze Tracker to the patient’s eye. It is imperative, therefore,
that the patient maintain the same position during gaze initialization and testing. Deviations are
recorded and displayed on the test screen and on the printout.
Note: Some patients with small pupils, ptotic lids, interfering lashes, or strong prescriptions may not
be good candidates for gaze monitoring.
On the Start of Test screen, you can make changes to the fixation monitoring system, or turn the
monitoring system off entirely by pressing CHANGE PARAMETERS and selecting the desired option.
You may select Gaze Tracking only at the start of a test; however, it may be turned off at any time
during
the test.
The Gaze Graph
The gaze graph is a useful method for documenting movement of the patient’s test eye. A test starts
with no markings on the gaze graph. As time progresses, the graph expands from the right, marking
eye movement and blinks.
Upward markings indicate that the test eye deviated from the fixation target at the time of stimulus
presentation. The higher the marking, the greater the deviation. The direction of deviation from the
fixation target is not indicated. Only the magnitude is recorded.
Содержание Humphrey II-i series
Страница 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Страница 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Страница 40: ...Introduction Instrument Setup Humphrey Field Analyzer II i series User Manual 2660021145640 A 1 32...
Страница 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Страница 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Страница 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Страница 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Страница 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Страница 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Страница 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Страница 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Страница 540: ...Installing Licensing HFA II i Software Humphrey Field Analyzer II i series User Manual 2660021145640 A J 10...
Страница 558: ...Reference to Older Test Strategies Humphrey Field Analyzer II i series User Manual 2660021145640 A L 10...
Страница 579: ......