Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
Networking
14-59
Imported Patient Rules and Conflicts
Patients imported into your HFA II-
i
may cause conflicts with your existing HFA II-
i
patient database
under certain conditions.
1. If the patient does not have a Patient ID:
A unique 29-character Patient ID such as “1966.1207.786F.C555.B6B9.473F” will be created for the
patient from the patient’s name and date of birth, and an HFA specific Issuer of ID will be entered
for the patient (1.2.276.0.75.2.2.30.2).
2. If the patient has a Patient ID and Issuer of ID (empty or defined):
1. If no other patient exists with this Patient ID and Issuer of ID, the patient will be saved as-is.
2. If a patient exists with the same Patient ID, Issuer of ID, names, and date of birth, then the
incoming patient will update the existing patient; the imported patient’s exams merges with
the existing patient.
WARNING: Every patient has an identifier that consists of Name, Date of Birth, Patient
ID, and Issuer of ID. To avoid unintentional automatic merging of patient records,
VERIFY a unique combination of Name, Date of Birth, Patient ID, and Issuer of ID
exists when patients are copied, transferred, retrieved from an archive, or restored
from a backup. There is no way to separate the combined records of erroneously
merged patients. Automatic merging of patient records occurs when a patient exists
in the local database with the EXACT SAME Name, Date of Birth, Patient ID, and
Issuer of ID as a patient that is copied, transferred, retrieved from an archive, or
restored from a backup.
For example, consider the following scenario:
Patient A: This patient is currently on the local HFA database and was automatically
assigned a unique Patient ID by the instrument because it was a blank field prior to a
software upgrade to V5.x software or during an import to the HFA through a copy,
transfer, retrieval from archive, or a restore from backup.
Patient B: This patient is about to be transferred to the local HFA database and has
the same Name and Date of Birth as Patient A, but does not have a Patient ID (i.e.
field is blank). When Patient B is copied, transferred, retrieved from an archive, or
restored from a backup to the local HFA database, Patient B will automatically be
assigned the same Patient ID and Issuer of ID as Patient A, and Patient B will be
merged with Patient A.
3.If a patient exists with the same Patient ID and Issuer of ID, but with a different name and/or
date of birth, a new Patient ID is created for the patient in the form “_ID<Old ID><New
ID>”, where <New ID> is created from the name and date of birth of the incoming patient as
described in
1. If the patient does not have a Patient ID:
of ID will be entered for the patient (1.2.276.0.75.2.2.30.2) and:
• If a patient exists in the database with this new Patient ID/Issuer of ID, the patient will be
saved with this new Patient ID/Issuer of ID.
Содержание Humphrey II-i series
Страница 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Страница 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Страница 40: ...Introduction Instrument Setup Humphrey Field Analyzer II i series User Manual 2660021145640 A 1 32...
Страница 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Страница 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Страница 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Страница 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Страница 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Страница 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Страница 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Страница 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Страница 540: ...Installing Licensing HFA II i Software Humphrey Field Analyzer II i series User Manual 2660021145640 A J 10...
Страница 558: ...Reference to Older Test Strategies Humphrey Field Analyzer II i series User Manual 2660021145640 A L 10...
Страница 579: ......