Care and Cleaning
Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
15-12
Alternative Touch Screen Calibration Method Number One
If you cannot press the
SYSTEM SETUP
icon on the Main Menu and you have an external keyboard
attached, use the following steps:
1
Press the F6 key.
2
Press the external keyboard’s Tab key until the ADDITIONAL SETUP button is highlighted.
3
Press the Enter/Return key on the external keyboard. Again press the Tab key until TOUCH
SCREEN CALIBRATION is highlighted.
4
Press the Enter/Return key again. Follow the instructions on the screen to continue the
standard calibration method as described starting with Step 5 on page
Alternative Touch Screen Calibration Method Number Two
The other method for touch screen calibration can be achieved when turning the HFA II-
i
power on.
1
If you continuously hold the patient response button down while the HFA II-
i
is powering up,
the option to calibrate the touch screen will be displayed. Calibrate the screen in the standard
calibration method as described starting with Step 5 on page
Note: If you accidentally get the touch screen calibration screen while starting up the HFA II-
i
, you
can continue the regular start up by cancelling out of the calibration mode. This situation
occasionally occurs when the patient response button is pressed down continuously because of the
way it was placed in its holder. Always make sure that the response button is not compressed when
it is placed in the holder.
Using Removable USB Storage Devices and Floppy Disks
All HFA II-
i
models can use removable USB storage devices and 1.44 MB 3.5" high density floppy
disks with an optional USB floppy disk drive. See
“Removable USB Storage Devices and USB Floppy
.
To ensure the integrity of the data, learn how to care for and handle these removable media
properly. The information on removable media can be destroyed by static electricity and strong
electromagnets. Typical sources of magnetic fields include telephones, fluorescent desk lamps,
magnetic desk accessories and other electrical appliances.
• Keep your removable media at least 5 feet from these sources of magnetic fields.
• Do NOT touch the recording surface of a floppy disk.
• Remember to label removable media for easy identification. Be careful to not let the label
adhere to the sliding cover of a floppy disk.
• Store floppy disks in their protective folders and in the original boxes supplied by the
manufacturer or in any filing system designed for the disks.
• Do NOT store floppy disks on top of the HFA II-
i
.
Содержание Humphrey II-i series
Страница 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Страница 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Страница 40: ...Introduction Instrument Setup Humphrey Field Analyzer II i series User Manual 2660021145640 A 1 32...
Страница 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Страница 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Страница 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Страница 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Страница 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Страница 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Страница 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Страница 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Страница 540: ...Installing Licensing HFA II i Software Humphrey Field Analyzer II i series User Manual 2660021145640 A J 10...
Страница 558: ...Reference to Older Test Strategies Humphrey Field Analyzer II i series User Manual 2660021145640 A L 10...
Страница 579: ......