38
Lamp replacement and calibration - Corona extreme | Service Manual | EN_40_070_0011I
Replacing the halogen lamp
Lamp Replacement and Calibration Corona extreme
2
Insert the base plate into the housing and secure
First check the correct mounting position. The electrical connectors of the
board and the housing need to be aligned to each other.
Caution
RISK OF COMPONENT DAMAGE
When the base plate is put back onto the housing, there is a risk of damage
to the boards, e.g. through individual components shearing off.
Therefore put the base plate very slowly and carefully back onto the
housing, visually monitoring the movement at all times.
[ A ]
Carefully put the base plate
1
in a straight line back onto the
housing
2
.
The two plug connections between the boards of the housing and the
base plate are reconnected.
[ B ]
Remove the 8 mounting pins
3
.
[ C ]
Screw the 12 fixing screws
4
back into the housing.
Use a TX 25 Torx screwdriver for this.
Fig. 42
3
A
B
C
4
1
2
Inserting the base plate into the housing and screw tight