Lamp replacement and calibration - Corona extreme | Service Manual | EN_40_070_0011I
Performing firmware update
Lamp Replacement and Calibration Corona extreme
9
Create *.ini file
Click on
Export details
under “Device info”.
An *.ini file is created.
Save this file.
Fig. 6 Creation of *.ini file
3 Performing firmware update
Note
IMPORTANT NOTE REGARDING FIRMWARE UPDATE
Do not switch off the device or interrupt the network connection tor the test
computer during the update under any circumstances.
Note
The current firmware can be found on the Internet at
http://www.zeiss.com/