Replacing the halogen lamp
Lamp Replacement and Calibration Corona extreme
3
Remove the lamp holder
Unscrew the two Torx screws
1
from the lamp holder. Use a TX 25 Torx
screwdriver for this.
Remove the entire lamp holder – consisting of the upper lamp holder ring
3
and lower lamp holder ring
4
together with the old halogen lamp
2
–
carefully upwards.
Place the lamp holder on a clean surface.
Fig. 22
1
3
2
4
Removing the lamp holder
4
Replace the halogen lamp
Lift the upper lamp holder ring
1
off the lower lamp holder ring
4
.
Put the gloves on.
Replace the old halogen lamp
3
with a new one from Maintenance
Package 1.
Replace the O-ring with a new one
2
after each lamp replacement.
Place the upper lamp holder ring back on the lower lamp holder ring.
Ensure that the holes match!
Fig. 23
2
3
4
1
Replacing the halogen lamp
NOTE:
Some type A lamps are installed
without the lower lamp holder
ring.
In this case, the lamp is replaced
as described here, but without
the lower lamp holder ring.