15
EN_40_070_0011I | Service Manual | Lamp replacement and calibration - Corona extreme
Performing firmware update
Lamp Replacement and Calibration Corona extreme
5
Run batch file (only Hardware Revision 2)
Note
IMPORTANT NOTE REGARDING FIRMWARE UPDATE
In order to update the device from revision 2 to revision 3, a corresponding
batch file needs to be run.
Download the folder “BatchFile_Rev2_to_Rev3” to your test computer.
Open the folder. It contains several files and subfolders. Do not remove these.
Use only the file “Install_Rev2_to_Rev3.bat”.
Start the batch file by double-clicking on
Install_Rev2_to_Rev3.bat
.
Read the info text.
Fig. 14 Confirmation prompt for batch file
Select
[y]
(Yes) if Hardware Revision 2 is installed and you intend to replace the
lamp.
Select
[n]
(No) if you do not intend to replace the lamp.
[y]
[n]