EN_40_070_0011I | Service Manual | Lamp replacement and calibration - Corona extreme
Replacing the halogen lamp
Lamp Replacement and Calibration Corona extreme
5
Install adapter cable for type B lamp
If necessary, open the right cable clip
1
. If necessary, use a slot-head
screwdriver for this.
Insert the adapter cable
2
into the cable clip and close it again.
Insert the wide plug of the adapter cable into connector / jack
3
on the lamp
board.
Fig. 28
3
1
2
Installing adapter cable for type B lamp
Caution
SENSITIVE OPTICAL COMPONENTS
Do not touch the reflector or the glass bulb of the new halogen lamp
without gloves.
Work on optical components in a very clean working enviroment only.
6
Insert new reflector and new O-ring
Insert the new reflector
2
into the ring groove of the lamp housing
1
.
Place the new O-ring
3
concentrically into the reflector
2
.
Fig. 29
3
2
1
Inserting new O-ring and new reflector