EN_40_070_0011I | Service Manual | Lamp replacement and calibration - Corona extreme
Preparing lamp replacement
Lamp Replacement and Calibration Corona extreme
1 Preparing lamp replacement
•
Determine the hardware revision status of the Corona extreme measuring
system
If possible, ask the customer for the serial number of the device.
You can obtain the hardware revision status of the device from the ZEISS
Service department ([email protected]) by giving the serial
number.
•
Download the Test Center Software
Make sure that the latest version of the Test Center Software is installed on
your test computer. This can be downloaded from tbd.
•
Have the following service / maintenance parts ready / on hand.
– Maintenance package
ȩ
Hardware revision 1: Maintenance Package 1,
order no. 000000-2397-831
ȩ
Hardware revision 2: Maintenance Package 2,
order no. 000000-2397-832
ȩ
Hardware revision 3: Maintenance Package 3,
order no. 000000-2402-119
– Ethernet cable, order no. 000000-2036-573
– Laboratory power supply unit, order no. 000000-2045-548
•
Check the white, gray and black reference standards
Caution
SENSITIVE SURFACES
Do not touch the surface of the white reference standard.
Make sure that the white, gray and black reference standards are clean. If
necessary, clean the reference standards with compressed air.
The white and gray reference standards are set in glass and can be cleaned
using alcohol.
Note
Make sure that the white and gray reference standard certificates are not
expired.
The corresponding certificate can be found on the USB memory stick included
in the delivery package.
Requirements