EN_40_070_0011I | Service Manual | Lamp replacement and calibration - Corona extreme
Performing sensor calibration
Lamp Replacement and Calibration Corona extreme
11
Photometric verification – Gray reference standard measurement
Place the gray reference standard on the centering aid.
Ensure correct alignment (interlocking of groove and pin).
Click
OK
to start the measurement.
Fig. 52 Calibrating gray reference standard
Remove the gray reference standard when the measurement is complete.
12
Photometric verification – White reference standard measurement
Place the white reference standard on the centering aid. Ensure correct
alignment (interlocking of groove and pin).
Click
OK
to start the measurement.
Fig. 53 Calibrating white reference standard