EN_40_070_0011I | Service Manual | Lamp replacement and calibration - Corona extreme
Performing sensor calibration
Lamp Replacement and Calibration Corona extreme
16
Back to Profile selection
Return to the profile selection to perform further profile configurations.
Fig. 60 Back to profile selection
17
Select a further test block / device profile
A scan is performed for potential / available test blocks / device profiles.
Select the next test block / device profile.
Fig. 61 Selecting a further device profile
Click
to proceed to the next step.
Profile:
Sapphire Flange
Reference holder:
“1462-668”
Profile:
Sample pan - 5.5 mm
thick base (in TURNSTEP)
Reference holder:
“2130-893”
Profile:
Sample pan -
thin base (in TURNSTEP)
Reference holder:
“2285-627”