27
EN_40_070_0011I | Service Manual | Lamp replacement and calibration - Corona extreme
Replacing the halogen lamp
Lamp Replacement and Calibration Corona extreme
4.4 Replacing halogen lamp – Hardware Revision 2
•
The base plate has been removed and placed on an ESD base
(see section “4.2 Opening the housing” on page 20).
•
If you have not already done so, put the following spare parts ready:
–
Maintenance package 2
, order no. 000000-2397-832
1
Disconnect the plug connections
Disconnect the lamp cable
1a
from the halogen lamp
2
.
Disconnect the lamp cable plug
1b
from the connector / jack
4
on the lamp
board. If necessary, use a slot-head screwdriver to unscrew the connection.
Proceed very carefully.
Fig. 26
1a
2
4
1b
3
Removing the old lamp cable
2
Remove the lamp cable
Open the two cable clips
3
. If necessary, use a slot-head screwdriver for this.
Proceed very carefully.
Remove the old lamp cable
1
from the cable clips.
Requirements
Procedure
NOTE:
Some type A lamps are installed
without the lower lamp holder
ring.
In this case, the lamp is replaced
as described here, but without
the lower lamp holder ring.
Typ