6-8
IM 704210-01E
Selecting the slope (edge)
Select the desired slope from the following. This is required only if EXT has been selected as the
arming source.
•
: Causes arming at a rising edge of the external arming signal.
•
: Causes arming at a falling edge of the external arming signal.
Selecting the arming delay type
Select the desired arming delay type from the following. This setting is available only when EXT
arming is selected.
• OFF : Selection/setting of arming delay is disabled. Arming delay will not be performed.
• TIME : Time delay
• EVENT : Event delay
• Setting the delay time
The delay time can be set within the following range. This setting is available only when TIME
arming delay is selected.
1
µ
s to 1 s
Note
Setting resolution varies depending on the value to be set. For details, refer to 12.2.
• Selecting the number of events to be delayed
Select from the following. This setting is available only when EVENT arming delay is selected.
1/2/3/4/5/6/7/8/9/10/20/30/40/50/60/70/80/90/100/200/300/400/500/600/700/800/900/1000/2000/
3000/4000/5000/6000/7000/8000/9000/10000/20000/30000
Note
The number of events to be delayed for the signal which is input to the measuring channel must be selected.
However, for time interval measurement, the number of events to be delayed for the signal input to the
channel A (CHA) must be selected.