12-9
IM 704210-01E
Specifications
12
12.10
General Specifications
Item
Specifications
Standard operating conditions
• Ambient temperature: 23
±
2ºC
• Ambient humidity: 50
±
10% RH
• Supply voltage/frequency: Within 1% of the rated voltage/frequency
Warm-up time
Approx. 30 minutes
Storage environment
• Temperature: –20 to +60ºC
• Humidity: 20 to 80% RH (no condensation allowed)
Operating environment
• Temperature: 5 to +40ºC
• Humidity: 20 to 80% RH (no condensation allowed)
Rated supply voltage
100 to 240 VAC
Permitted supply voltage range
90 to 264 VAC
Rated supply voltage frequency
50 to 60 Hz
Permitted supply voltage frequency range
48 to 63 Hz
Power consumption:
125 VA max.
Withstand voltage (between power supply and case)
1.5 kVAC, for 1 minute
Insulation resistance (between power supply and case)
10M
Ω
min, at 500 VDC.
Signal ground
All the input/output connectors must be grounded to the case.
External dimensions: Approx.
213 (W)
×
132 (H)
×
350 (D) mm, excluding projections
Weight
Approx. 5 kg (main unit only)
Cooling method
Forced air cooling
Installation position
Horizontal (Placing instruments one on top of the other is not allowed.)
Battery back-up
Set-up information and clock data are backed-up by internal lithium batteries.
Standard accessories
• Power cord: 1
• Rubber foot: 2
• User’s Manual: 1 (this manual)
• GP-IB Interface User’s Manual: 1