![YOKOGAWA TA320 Скачать руководство пользователя страница 135](http://html.mh-extra.com/html/yokogawa/ta320/ta320_user-manual_911358135.webp)
12-4
IM 704210-01E
12.3 Gate, Arming, Inhibit
Item
Specifications
Gating
• Type: Can be selected from EVENT, TIME and EXTERNAL.
• Configurable number of events (within 320 s following the arming)
• Time stamp mode
• Period, time interval, pulse width measurements: 1 to 32,000
• Phase difference, duty ratio: 1 to 16,000
• Hardware histogram mode: 1 to 99,999,999
• Configurable gate time range (within maximum number of events for sampling mode)
• 1
µ
s
gate time < 5 ms (in 200 ns increments)
• 5 ms
gate time
10 s (in 1 ms increments)
• Allowable time frame for external gate
1
µ
s to 320 s (within maximum number of events for sampling mode)
• Polarity: Can be selected from
and
.
Arming
• Arming source: Can be selected from AUTO (automatic) and EXT (external).
• Arming delay (only when EXT is selected as the arming source)
• Configurable delay time in the case of time delay
• 1
µ
s
delay time < 5 ms (in 200 ns increments)
• 5 ms
dalay time
1 s (in 1 ms increments)
• Event delay: Can be set from 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 20, 30, 40, 50, 60, 70, 80, 90, 100, 200, 300,
400, 500, 600, 700, 800, 900, 1000, 2000, 3000, 4000, 5000, 6000, 7000, 8000, 9000, 10000, 20000,
and 30000.
• Slope: Can be selected from
and
.
Inhibit
• Allowable time frame: 1
µ
s to 320 s
• Polarity: Can be selected from
and
.