![YOKOGAWA TA320 Скачать руководство пользователя страница 137](http://html.mh-extra.com/html/yokogawa/ta320/ta320_user-manual_911358137.webp)
12-6
IM 704210-01E
Item
Specifications
Time variation display (time stamp mode only)
• Scale: X-axis and Y-axis of time variation
• X MIN (minimum value of X-axis): 0 to 199.9999800 sec.
• X SPAN (maximum value of X-axis): Can be selected from 20, 40, 100, 200, 400
µ
s, 1, 2, 4, 10, 20,
40, 100, 200, 400 ms, 1, 2, 4, 10, 20, 40, 100 and 200 sec.
• Y CENTER (center value of Y-axis)
• Period, time interval and pulse width measurements: 10 ns to 99.9999900 ms
• Phase difference measurement: 15.00 to 345.00 deg
• Duty ratio measurement: 0.500 to 99.500%
• Y SPAN (span of Y-axis)
• Period, time interval and pulse width measurements: Can be selected from 2, 4, 10, 20, 40, 100, 200,
400 ns, 1, 2, 4, 10, 20, 40, 100, 200, 400
µ
s, 1, 2, 4, 10, 20, 40 and 100 ms.
• Phase difference measurement: Can be selected from 30, 60, 90, 180 and 360 deg.
• Duty ratio measurement: Can be selected from 1, 5, 10, 50 and 100%.
• Readout: Markers can be used to read measured values present in the area enclosed by the two markers.
• Display parameters: Grid and interpolation displays can be turned ON and OFF.
• X-axis resolution (time stamp mode): 100 ns
List display (time stamp mode)
A list of time stamps and their measured values is displayed.
• Displayed data can be scrolled up and down.
• Resolution for time stamps: 100 ns
List display (hardware histogram mode)
A list of measured values (class values of histogram) and their frequency is displayed.
Displayed data can be scrolled up and down.
Statistics display (time stamp mode)
A list of statistic values for the area specified for histogram display calculation setting is displayed.
Statistical computation items *: Maximum value, minimum value, mean value, standard deviation, peak to
peak, flutter
Statistics display (hardware histogram mode)
A list of statistic values for the area specified for histogram display calculation setting is displayed.
• Statistical computation items *: Maximum value, minimum value, mean value, standard deviation, peak
to peak, flutter, jitter 1, jitter 2
• Standard deviation is displayed in bar graph.
Bar graph range: Can be selected from 7.5, 15, 30 and 60 ns.
*: Statistic values in the case of
selected as the polarity for pulse width measurement with , and
selected as the polarity for
the calculation menu.
• Maximum, minimum, average, peak to peak, flutter, jitter 2: Not computed.
• Standard deviation, jitter 1: Standard deviation and jitter 1 are computed, individually for positive and negative sides, then the sum
of computed standard deviation and that of computed jitter 1 are divided by 2.
12.4 Display