11-2
IM 704210-01E
11.2 Messages
The following three kinds of message may appear on the screen.
• Error messages
Displayed when an attempt is made to execute an operation which is not executable or when an
abnormality occurs with the instrument.
• Warning messages
Displayed when the instrument is in such a state that the user needs to be warned (e.g. when the
measurement data overflows).
• Status messages
Displayed to indicate the current state of the instrument. They are displayed at the bottom of
the screen. (e.g. AUTO LOADING!)
This section describes the cause of each error message (except for communications related error
messages) and the corrective action to be taken, in addition to the meaning of each warning
message. For communications related error messages, refer to the GP-IB Interface User’s Manual
(IM704210-12E).
Error Messages
Error messages (except for those related to communications) are displayed in the center of the
screen.
* Error messages related to communications are displayed in the error log window, like warning
messages.
No.
Message
Reference Section
602
No disk inserted.
9.1
No floppy disk is inserted in the floppy disk drive. Insert an appropriate disk.
604
Floppy disk isn’t formatted.
9.2
Floppy disk has not been formatted. Format it.
605
File not found.
–
File does not exist. Check the file.
606
Disk is write-protected.
–
Floppy disk is write-protected. Remove write-protection.
609
Same file already exists.
–
A file with the same file name exists. Assign another name to the file and save it.
610
InValid file name.
9.3, 9.4, 9.5
The filename is incorrect. Enter the correct name.
612
Disk is full.
9.2
Floppy disk has insufficient space. Use a new floppy disk.
615
Disk format failure.
9.2
Formatting of the disk failed. Format it again.
620
File access failure.
–
An abnormality has occurred during access to the file. The file is probably damaged.
621
InValid data format.
–
Data format is invalid. Check the file.
671
Measurement is in progress.
9.3, 9.4
It is not possible to access a file during measurement.
672
Data to be saved don’t exist.
–
Data to be saved does not exist.
701
During output to plotter.
10.1
It is not possible to operate the instrument during output to the plotter.
906
Fan stopped. Turn OFF the power immediately.
–
Cooling fan has stopped. Turn off the power immediately.