12-1
IM 704210-01E
Specifications
12
12.1 Measuring Inputs and Trigger
Item
Specification
Number of input channels
2 (CHA and CHB)
Input coupling
DC/AC
Input connector
BNC connector
Input impedance
50
Ω
/1 M
Ω
, 20 pF (typical *
1
)
Frequency characteristic (-3 dB) • DC input coupling: DC to 100 MHz (typical *
1
)
• AC input coupling
• 680 KHz to 100 MHz (typical *
1
) (Input impedance 50
Ω
)
• 35 Hz to 100 MHz (typical *
1
) (Input impedance 1 M
Ω
)
Internal jitter
300 ps (rms)
Minimum input pulse width
10 ns
Operating voltage range
–5 V to +5 V
Maximum input voltage
• Input impedance 50
Ω
: 5V
rms
• Input impedance 1 M
Ω
• 40 V (DC + AC
peak
) (DC
input frequency
100 KHz)
• {3.5/f (MHz) + 5} V (DC + AC
peak
) (100 kHz < input frequency
100MHz)
Input sensitivity *
2
100 mV
p-p
Input amplifier noise
400
µ
V
rms
(typical *
1
)
Trigger
• Trigger mode : Can be selected from single auto trigger, repeat auto trigger and manual trigger
• Trigger level
• Setting range: –5 V to +5 V (can be set using front panel keys for manual trigger mode)
• Accuracy *
2
:
±
(10 mV + 4% of predetermined value)
• Resolution: 5 mV
• Input signal condition: 1 kHz to 50 MHz (single auto trigger, repeat auto trigger)
• Setpoint time: 0.7 s (typical *1) (single auto trigger, repeat auto trigger)
Sampling
• Sampling mode: Can be selected from time stamp mode and hardware histogram mode
• Maximum sampling rate:
Continuous 14 M samples/s (approx. 71-ns intervals)
7 M samples/s (approx. 142-ns intervals) in the case of duty ratio and phase difference measurements
• Maximum sample size (maximum number of data items)
• Time stamp mode: 32,000
(16,000 in the case of duty ratio and phase difference measurements)
• Hardware histogram mode: 99,999,999
(excluding duty ratio and phase difference measurements)
• Sampling interval
MIN, 1
µ
s, 2
µ
s, 4
µ
s, 10
µ
s, 20
µ
s, 40
µ
s, 100
µ
s, 200
µ
s, 400
µ
s, 1 ms, 2 ms, 4 ms
MIN is equivalent to the interval for the maximum sampling rate.
Must be selected for time stamp mode only.
• Maximum sampling time: 320 s (after the arming)
*1: Typical values are described in terms of typical of average performance. However, they cannot be guaranteed.
*2: Measured under the reference operating conditions specified in 12.10 “General Specifications”, after elapse of warm-up time, with the
DC input coupling and 1M
Ω
input impedance.