12-2
IM 704210-01E
12.2 Measurement Functions (Measurement Items)
Items
Specifications
Measurement Update Rate *
1
400 ms (hardware histogram mode only)
(For measurement of period of 1-MHz sine-wave, with sample size of 1000)
Period
• Measuring range
• 30 ns to 100 ms (time stamp mode)
• 30 ns to 3.2
µ
s (hardware histogram mode)
• Display resolution
• 100 ps (time stamp mode)
• 100 ps or histogram span setting/200, whichever larger (hardware histogram mode)
• Measurement resolution *
1
•
±
300 ps rms
±
2
×
trigger error *
2
(time stamp mode)
•
±
300 ps rms or display resolution
±
2
×
trigger error *
2
, whichever larger (hardware histogram mode)
• Accuracy *
1
±
300 ps rms
±
2
×
trigger error *
2
±
(timebase frequency stability
×
measured period)
±
300 ps
systematic error
• Slope: Can be selected from
and
.
Time Interval
• A-to-B Interval
• Measuring range
• 5 ns to 100 ms (time stamp mode)
• 5 ns to 3.2
µ
s (hardware histogram mode)
• Display resolution
• 100 ps (time stamp mode)
• 100 ps or histogram span setting/200, whichever larger (hardware histogram mode)
• Measurement resolution *
1
• Time stamp mode
• When slope is
A
B
/
A
B
/
A
B
/
A
B
±
300 ps rms
±
A-input trigger error *
2
±
B-input trigger error *
2
• When slope is
A
B
/
A
B
±
300 ps rms
±
A-input trigger error *
2
±
B-input trigger error *
2
±
trigger level timing error *
3
• Hardware histogram mode
• When slope is
A
B
/
A
B
/
A
B
/
A
B
±
300 ps rms or display resolution, whichever larger
±
A-input trigger error *
2
±
B-input trigger error *
2
• When slope is
A
B
/
A
B
±
300 ps rms or display resolution, whichever larger
±
A-input trigger error *
2
±
B-input trigger error *
2
±
trigger level timing error *
3
• Accuracy *
1
±
300 ps rms
±
trigger level timing error *
3
±
(timebase frequency stability
×
measured interval)
±
1-ns
systematic error
• Slope: Seletable from
A
B
/
A
B
/
A
B
/
A
B
/
A
B
/
A
B
.
• Condition for continuous measurement *
1
: Intervals between A-to-B measurements must be 30 ns or more
and the period of CHA signal must be 71 ns or more.
• A-to-B-to-A Interval
• Measuring range
• A to B: 30 ns to 100 ms, B to A: 30 ns to 100 ms (time stamp mode)
• A to B: 30 ns to 3.2
µ
s, B to A: 30 ns to 3.2
µ
s (hardware histogram mode)
• Display resolution
• 100 ps (time stamp mode)
• 100 ps or histogram span setting/200, whichever larger (hardware histogram mode)
• Measurement resolution *
1
•
±
300 ps rms
±
A-input trigger error *
2
±
B-input trigger error *
2
(time stamp mode)
•
±
300 ps rms or display resolution, whichever larger
±
A-input trigger error *
2
±
B-input timing error *
2
(hardware histogram mode)
• Accuracy *
1
±
300 ps rms
±
trigger level timing error *
3
±
(timebase frequency stability
×
measured interval)
±
1-ns
systematic error
• Slope: Selectable from
A B A
/
A B A
/
A B A
/
A B A
.