![YOKOGAWA TA320 Скачать руководство пользователя страница 23](http://html.mh-extra.com/html/yokogawa/ta320/ta320_user-manual_911358023.webp)
1-13
IM 704210-01E
Functions
1
Panorama Display
≡
For operating procedure, refer to 7.3
≡
A panorama bar is displayed at the top of the screen during hardware histogram mode. The
panorama bar shows all the histograms of measured data. The histogram currently displayed in
the window is enclosed by “ marks,” so that the user can easily understand which histogram
is currently displayed.
Center value of X-axis
(Class value of
the center class)
Right-end value of X-axis
(positive displacement
from the center value of X-axis)
Span of X-axis
(40 ns in this example)
Left-end value of X-axis
(negative displacement from
the center value of X-axis)
Maximum value of Y-axis
Y-axis
(frequency)
Panorama bar
Window 1
Window 2
Window 3
Window 4
Window
(Window 1 is displayed
in this example)
Left-end value of
the panorama bar
Right-end value of
the panorama bar
mark
X-axis (time)
Histogram Data Addition
≡
For operating procedure, refer to 7.4
≡
It is possible to add the histogram of each window specified using the SIZE touch key during
multi-window setting, and display the resulting histogram in a single window. The frequency of
the bin located at the center value of X-axis of each histogram is added, and then the frequency of
each bin located on the left and right of that bin is similarly added. This enables comparison of
data distribution between histograms and global statistical computation of all the histograms.
+
=
Histogram for
window 1
Histogram for
window 2
Resulting histogram
Time Variation Display
≡
For operating procedure, refer to 7.5
≡
During time stamp mode, measured values and time stamps are treated as data and acquired in the
acquisition memory. The time variation display function uses this time-series data to display
variation of measured values in relation to the elapsed time following the arming.
Maximum value of X-axis
(positive displacement from
the minimum value of X-axis)
(Span: 400
µ
s in this example)
Minimum value
of X-axis
X-axis (elapsed time)
Center value
of Y-axis
Span of Y-axis
(20 ns in this example)
(negative displacement from the
center value of Y-axis)
(positive displacement from the
center value of Y-axis)
Y-axis
(measured value)
1.6 Displaying the Measured/Computed Data