9-6
IM 704210-01E
Explanation
Set-up information to be saved
Settings made and selected using touch keys, rotary knob and ten-key are saved. In addition, the
set-up information stored in the internal memory is also saved in the auto load file.
File name extension and directory
An extension (.SET) is attached to the file name automatically, and the file is saved in the
directory TA_SETUP. The auto load file is called AUTOLOAD.SET and saved in the root
directory.
Data size
Each set-up information requires approximately 8 KB. The auto load file can contain data of
approximately 50 KB in total.
Checking the free space available on a floppy disk
Refer to 9.2.
Entering the file name
• The file name must consist of 8 or less characters.
• The following characters can be used for file name.
Alphabets (capital letter), numerical values, – (minus), _ (under-bar)
• The following file names cannot be used due to restrictions on MS-DOS.
AUX, CON, PRN, NUL, CLOCK
• Saving under a file name already existing in the same directory is not allowed. (Overwriting is
not allowed.)
Note
• Loading is not possible during measurement (i.e. while the START indicator is ON).
• If the extension of a file name is changed using a personal computer etc., it will be impossible to load that
file.
CAUTION
Never remove the floppy disk while the access indicator is blinking. Failure to do so
may damage the magnetic head of the floppy disk drive or destroy the data saved on the
floppy disk.
9.3 Saving, Loading and Auto Loading the Set-up Information