7-2
IM 704210-01E
Setting the X-axis and Y-axis
5.
Press the SCALE touch key to display the X-axis/Y-axis menu.
• Setting the center value of X-axis
6.
Press the X CENTER touch key to highlight X CENTER.
7.
Turn the rotary key or use the ten-key to set the desired center value of X-axis.
• Selecting the span of X-axis
8.
Press the X SPAN touch key to highlight X SPAN.
9.
Turn the rotary key to select the desired span of X-axis.
• Selecting the Y-axis scale
10. Press the Y AXIS touch key to highlight LIN or LOG. The highlighted scale will be
selected.
• Selecting the maximum value of Y-axis
11. Press the Y HIGH touch key to highlight Y HIGH.
12. Turn the rotary knob to select the desired maximum value of Y-axis.
• Returning to the DISPLAY top menu
13. Press the ESC touch key.
Setting readout parameters
14. From the DISPLAY top menu, press the MARKER touch key to display the marker menu.
• Turning marker display ON/OFF
15. Press the MARKER touch key of the marker menu, to highlight OFF or ON. The
highlighted option (OFF or ON) will be selected. If OFF is selected, skip to step 23.
• Moving the X1 marker
16. Press the X1 MARKER touch key to highlight X1 MARKER.
17. Turn the rotary knob to move the X1 marker to the desired position.
• Moving the X2 marker
18. Press the X2 MARKER touch key to highlight X2 MARKER.
19. Turn the rotary knob to move the X2 marker to the desired position.
• Selecting the marker speed
20. Press the MOVEMENT touch key to highlight
. This will cause the marker to move
faster.
21. Press the MOVEMENT touch key to highlight . This will cause the marker to move at the
original speed.
• Returning to the DISPLAY top menu
22. Press the ESC touch key.
Setting the calculation function
23. From the DISPLAY top menu, press the CALCULATION touch key to display the
calculation menu.
• Selecting the area
24. Press the AREA touch key to highlight AREA.
25. Turn the rotary knob to select the desired area.
• Setting constant T
26. Press the CONST T touch key to highlight CONST T.
27. Turn the rotary knob or use the ten-key to set the desired constant.
• Selecting the polarity to be used for display/statistical computation (Required only
if
has been selected for pulse width measurement)
28. Press the POLARITY touch key to highlight
,
or
. The highlighted polarity
will be selected.
X-axis/Y-axis menu
Sets the center
value of X-axis.
Sets the span
of X-axis.
Sets the maximum
value of Y-axis.
Sets the
Y-axis scale.
Displayed only when the
multi-window function is
active (see 7.2).
Marker menu
Turns marker
display ON/OFF.
Moves the
X1 marker.
Moves the
X2 marker.
Selects the
marker speed.
Calculation menu
Sets the
constant T.
Also used for statistical
computation (see 7.7 and 7.8).
Selects
the area.
Selects the
polarity to be
used for
display/statistical
computation.
Displayed only when
is selected for pulse width
measurement.
7.1 Setting the Histogram Display Function and Reading the Data (Readout)