7.2 Using the Multi-window Function
Setting the multi-window function
9.
Press the WINDOW touch key of the DISPLAY top menu to display the multi-window menu.
• Selecting multi-window
10. Press the SINGLE/MULTI touch key to highlight MULTI. The MODE and SIZE touch keys
will appear.
• Selecting the multi-window mode
11. Press the MODE touch key to highlight AUTO or MAN. The highlighted mode (AUTO or
MAN) will be selected.
• Selecting the window size (number of windows)
12. Press the SIZE touch key to highlight SIZE.
13. Turn the rotary knob to set the desired window size.
• Returning to the DISPLAY top menu
14. Press the ESC touch key to return to the DISPLAY top menu.
Multi-window menu
Selects the multi-window function.
Selects the multi-window mode.
Panorama display menu (see 7.3)
Selects the window size.
Setting the X-axis and Y-axis
15. Press the SCALE touch key to display the X-axis/Y-axis menu.
• Setting the center value of X-axis
16. Press the X CENTER touch key to highlight X CENTER.
17. Turn the rotary key or use the ten-key to set the desired center value of X-axis for window 1.
The value recorded at step 6 (the X-axis coordinate of the X1 marker) must be set.
• Selecting the span of X-axis
18. Press the X SPAN touch key to highlight X SPAN.
19. Turn the rotary key to select the desired span of X-axis. By referring to the value recorded at
step 7 (the displacement of the X2 marker from X1 marker), set the span so that the entire
histogram to be analyzed is positioned inside the window.
• Setting the space between the center values of X-axis (required only when AUTO
multi-window mode is selected)
20. Press the W.SPACE touch key to highlight W.SPACE.
21. Turn the rotary key or use the ten-key to set the space between the center value of X-axis for
a window and that for other windows. The value recorded at step 7 (the displacement of the
X2 marker from X1 marker) must be set.
• Selecting the scale and maximum value of Y-axis
22. Select the scale and maximum value of Y-axis, if necessary. For the selection method, refer to 7.1.
• Re-setting the X-axis and Y-axis
23. Repeat steps from 15 until the histogram can be observed clearly and easily.
X-axis/Y-axis menu
Sets the center value of X-axis.
Selects the span of X-axis.
Sets the space between the center values of X-axis.
(This setting is required only when AUTO
multi-window mode is selected.)
Selects the maximum value of Y-axis.
Selects the scale of Y-axis.