10.5 Setting the Auxiliary Trigger Source (AUX OUT)
Procedure
1.
Press
SHIFT+SETUP (MISC)
to display the MISC menu.
2.
Press the
External I/O
soft key to display the External I/O menu.
Setting the Auxiliary Trigger Output Source
3.
Press the
Aux Out
soft key to display the Auxiliary Out menu.
4.
Press the soft key corresponding to the desired auxiliary trigger output source.
Explanation
Select the auxiliary trigger output source from the four types below. For details on the
auxiliary trigger source, see “Trigger Block Diagram” in section 2.6, “Triggering” (page
2-27).
•
MeasBusy
Indicates measurement-in-progress. Transmits a low signal when the measurement
operation starts and a high signal when the measurement operation ends.
•
Timer1
Generates a 10-µs low-edge pulse when Timer1 is generated.
•
Timer2
Generates a 10-µs low-edge pulse when Timer2 is generated.
•
Through
Passes through the auxiliary input signal without any change.
<<Corresponding Command Mnemonic>>
:ROUTe:AUXiliary SENSe|TIMer1|TIMer2|THRough