App
Index
Log Sweep
(See section 6.2 for the procedure)
Sweeps the source level from the start level to the stop level in an exponential form
divided by the step count.
The step count can be set to any value in the range of 2 to 100000. However, if the start
value and stop value have opposite signs, an error occurs, and the sweep operation
cannot be carried out.
Step count
Stop level
Start level
When source waveform is DC
When source waveform is pulse
Stop level
Start level
Step count
Pulse base value
Trigger count
Source level
Sweep start
Source level
Trigger count
Sweep start
Trigger Trigger Trigger Trigger Trigger
Trigger Trigger Trigger Trigger Trigger
Start Level and Stop Level
• Setting Resolution of the Start level and Stop Level for Voltage Source Mode
765601/765602
|Start Level| and |Stop Level|
Setting Resolution
0.000 mV ≤ X ≤ 200.000 mV
1 µV
0.20000 V < X ≤ 2.00000 V
10 µV
2.0000 V
< X ≤ 7.0000 V
100 µV
7.0000 V
< X ≤ 18.0000 V
100 µV
765611/765612
|Start Level| and |Stop Level|
Setting Resolution
0.000 mV ≤ X ≤ 200.000 mV
1 µV
0.20000 V < X ≤ 2.00000 V
10 µV
2.0000 V
< X ≤ 20.0000 V
100 µV
20.0000 V < X ≤ 50.0000 V
100 µV
• Setting Resolution of the Start level and Stop Level for Current Source Mode
765601/765602
|Start Level| and |Stop Level|
Setting Resolution
0.000 nA
≤ X ≤ 200.000 nA
1 pA
0.20000 µA < X ≤ 2.00000 µA
10 pA
2.0000 µA
< X ≤ 20.0000 µA
100 pA
20.000 µA
< X ≤ 200.000 µA
1 nA
0.20000 mA < X ≤ 2.00000 mA
10 nA
2.0000 mA < X ≤ 20.0000 mA
100 nA
20.000 mA < X ≤ 200.000 mA
1 µA
0.20000 A
< X ≤ 3.20000 A
10 µA
765611/765612
|Start Level| and |Stop Level|
Setting Resolution
0.000 nA
≤ X ≤ 200.000 nA
1 pA
0.20000 µA < X ≤ 2.00000 µA
10 pA
2.0000 µA
< X ≤ 20.0000 µA
100 pA
20.000 µA
< X ≤ 200.000 µA
1 nA
0.20000 mA < X ≤ 2.00000 mA
10 nA
2.0000 mA < X ≤ 20.0000 mA
100 nA
20.000 mA < X ≤ 200.000 mA
1 µA
0.20000 A
< X ≤ 1.20000 A
10 µA
Step Count
Set a value in the range from 2 to 100000.
2.5 Sweep
2-21
IM 765601-01E
Explanation of Functions